PF series gas mass flow controllers

PF-5SLPM-D
Gas mass flow controller — 0 – 5 SLPM

PF-5SLPM-D is a gas mass flow controller covering 0 – 5 SLPM, with 1/8" NPT female process connections, 2.0 PSID pressure drop at full scale and an operating pressure range of 11.5 – 160 PSIA.

Request a quotation

Specification

ModelPF-5SLPM-D
TypeGas mass flow controller
Full-scale range0 – 5 SLPM
Process connection1/8" NPT female
Pressure drop (air to atmosphere, full scale)2.0 PSID
Operating pressure11.5 – 160 PSIA

What this range covers

bench-scale duty — laboratory reactors, cell-culture gas overlay, analytical instruments and gas-mixing panels where the working flow sits in the single-digit to twenty-litre band.

The published figure is 2.0 PSID measured with air exhausting to atmosphere at full scale. As a controller this figure includes the integral control valve, so the upstream supply must hold at least this differential above downstream at full-scale flow.

As a rule of thumb, size the instrument so your working flow falls between roughly 30% and 90% of full scale — well above the bottom of the range, where resolution is poor, and below the top, where the pressure drop figure quoted here applies. Instrument calibration is gas-specific: tell us the gas you are metering and we will confirm whether the published range applies directly or needs a correction.

Before you order — three things to confirm

  1. The gas. Calibration follows the gas you name. Reactive, corrosive or gas-mixture duty may need different wetted materials and seals — tell us the composition.
  2. The pressures either side. The pressure-drop column is measured to atmosphere at full scale. If your downstream pressure is above atmosphere, the usable span shifts; give us upstream and downstream pressure and we will confirm whether PF-5SLPM-D still fits.
  3. The connection. This unit is supplied with 1/8" NPT female. Check your mating fitting and tube size before ordering — an adaptor changes both the dead volume and the effective restriction on low-range units.

Other models in the PF series gas mass flow controllers

ModelTypeFull-scale rangeProcess connectionPressure dropOperating pressure
PF-1000SLPM-DGas mass flow controller0 – 1000 SLPM3/4" NPT female14.0 PSID11.5 – 160 PSIA
PF-100SCCM-DGas mass flow controller0 – 100 SCCM1/8" NPT female1.0 PSID11.5 – 160 PSIA
PF-100SLPM-DGas mass flow controller0 – 100 SLPM1/4" NPT female15.5 PSID11.5 – 160 PSIA
PF-10SLPM-DGas mass flow controller0 – 10 SLPM1/8" NPT female5.5 PSID11.5 – 160 PSIA
PF-1SLPM-DGas mass flow controller0 – 1 SLPM1/8" NPT female1.5 PSID11.5 – 160 PSIA
PF-200SLPM-DGas mass flow controller0 – 200 SLPM——11.5 – 160 PSIA
PF-20SLPM-DGas mass flow controller0 – 20 SLPM1/8" NPT female12.0 PSID11.5 – 160 PSIA
PF-2SLPM-DGas mass flow controller0 – 2 SLPM1/8" NPT female3.0 PSID11.5 – 160 PSIA
PF-500SCCM-DGas mass flow controller0 – 500 SCCM1/8" NPT female1.0 PSID11.5 – 160 PSIA
PF-500SLPM-DGas mass flow controller0 – 500 SLPM3/4" NPT female6.5 PSID11.5 – 160 PSIA

See the full PF series gas mass flow controllers table or compare all 37 ranges.