DONGDE YUNHAIAlicat Flow Instrument Specialists
PF series gas mass flow controllers

PF-100SCCM-D
Gas mass flow controller — 0 – 100 SCCM

PF-100SCCM-D is a gas mass flow controller covering 0 – 100 SCCM, with 1/8" NPT female process connections, 1.0 PSID pressure drop at full scale and an operating pressure range of 11.5 – 160 PSIA.

Request a quotation

Specification

ModelPF-100SCCM-D
TypeGas mass flow controller
Full-scale range0 – 100 SCCM
Process connection1/8" NPT female
Pressure drop (air to atmosphere, full scale)1.0 PSID
Operating pressure11.5 – 160 PSIA

What this range covers

sub-litre-per-minute duty — chromatography carrier and sample gas, leak and seal testing, small-batch gas blending, and bench rigs where a few hundred standard cubic centimetres per minute is the whole measurement.

The published figure is 1.0 PSID measured with air exhausting to atmosphere at full scale. As a controller this figure includes the integral control valve, so the upstream supply must hold at least this differential above downstream at full-scale flow.

As a rule of thumb, size the instrument so your working flow falls between roughly 30% and 90% of full scale — well above the bottom of the range, where resolution is poor, and below the top, where the pressure drop figure quoted here applies. Instrument calibration is gas-specific: tell us the gas you are metering and we will confirm whether the published range applies directly or needs a correction.

Before you order — three things to confirm

  1. The gas. Calibration follows the gas you name. Reactive, corrosive or gas-mixture duty may need different wetted materials and seals — tell us the composition.
  2. The pressures either side. The pressure-drop column is measured to atmosphere at full scale. If your downstream pressure is above atmosphere, the usable span shifts; give us upstream and downstream pressure and we will confirm whether PF-100SCCM-D still fits.
  3. The connection. This unit is supplied with 1/8" NPT female. Check your mating fitting and tube size before ordering — an adaptor changes both the dead volume and the effective restriction on low-range units.

Other models in the PF series gas mass flow controllers

ModelTypeFull-scale rangeProcess connectionPressure dropOperating pressure
PF-1000SLPM-DGas mass flow controller0 – 1000 SLPM3/4" NPT female14.0 PSID11.5 – 160 PSIA
PF-100SLPM-DGas mass flow controller0 – 100 SLPM1/4" NPT female15.5 PSID11.5 – 160 PSIA
PF-10SLPM-DGas mass flow controller0 – 10 SLPM1/8" NPT female5.5 PSID11.5 – 160 PSIA
PF-1SLPM-DGas mass flow controller0 – 1 SLPM1/8" NPT female1.5 PSID11.5 – 160 PSIA
PF-200SLPM-DGas mass flow controller0 – 200 SLPM——11.5 – 160 PSIA
PF-20SLPM-DGas mass flow controller0 – 20 SLPM1/8" NPT female12.0 PSID11.5 – 160 PSIA
PF-2SLPM-DGas mass flow controller0 – 2 SLPM1/8" NPT female3.0 PSID11.5 – 160 PSIA
PF-500SCCM-DGas mass flow controller0 – 500 SCCM1/8" NPT female1.0 PSID11.5 – 160 PSIA
PF-500SLPM-DGas mass flow controller0 – 500 SLPM3/4" NPT female6.5 PSID11.5 – 160 PSIA
PF-5SLPM-DGas mass flow controller0 – 5 SLPM1/8" NPT female2.0 PSID11.5 – 160 PSIA

See the full PF series gas mass flow controllers table or compare all 37 ranges.

Need this model quoted?

Send the gas, the flow you need to measure or control, and the pressures either side of the instrument. We will confirm the range and quote it — normally the same working day.

Request a quotation