DONGDE YUNHAIAlicat Flow Instrument Specialists
PFU / PFV corrosion-resistant controllers

PFV-5SLPM-D
Corrosion-resistant gas mass flow controller — 0 – 5 SLPM

PFV-5SLPM-D is a corrosion-resistant gas mass flow controller covering 0 – 5 SLPM, with 1/8" NPT female process connections, 2.0 PSID pressure drop at full scale and an operating pressure range of 11.5 – 160 PSIA.

Request a quotation

Specification

ModelPFV-5SLPM-D
TypeCorrosion-resistant gas mass flow controller
Full-scale range0 – 5 SLPM
Process connection1/8" NPT female
Pressure drop (air to atmosphere, full scale)2.0 PSID
Operating pressure11.5 – 160 PSIA

What this range covers

bench-scale duty — laboratory reactors, cell-culture gas overlay, analytical instruments and gas-mixing panels where the working flow sits in the single-digit to twenty-litre band.

The published figure is 2.0 PSID measured with air exhausting to atmosphere at full scale. As a controller this figure includes the integral control valve, so the upstream supply must hold at least this differential above downstream at full-scale flow.

As a rule of thumb, size the instrument so your working flow falls between roughly 30% and 90% of full scale — well above the bottom of the range, where resolution is poor, and below the top, where the pressure drop figure quoted here applies. Instrument calibration is gas-specific: tell us the gas you are metering and we will confirm whether the published range applies directly or needs a correction.

Before you order — three things to confirm

  1. The gas. Calibration follows the gas you name. Reactive, corrosive or gas-mixture duty may need different wetted materials and seals — tell us the composition.
  2. The pressures either side. The pressure-drop column is measured to atmosphere at full scale. If your downstream pressure is above atmosphere, the usable span shifts; give us upstream and downstream pressure and we will confirm whether PFV-5SLPM-D still fits.
  3. The connection. This unit is supplied with 1/8" NPT female. Check your mating fitting and tube size before ordering — an adaptor changes both the dead volume and the effective restriction on low-range units.

Other models in the PFU / PFV corrosion-resistant controllers

ModelTypeFull-scale rangeProcess connectionPressure dropOperating pressure
PFU-100SLPM-DCorrosion-resistant gas mass flow controller0 – 100 SLPM1/4" NPT female2.5 PSID11.5 – 160 PSIA
PFU-250SLPM-DCorrosion-resistant gas mass flow controller0 – 250 SLPM1/2" NPT female2.4 PSID11.5 – 160 PSIA

See the full PFU / PFV corrosion-resistant controllers table or compare all 37 ranges.

Need this model quoted?

Send the gas, the flow you need to measure or control, and the pressures either side of the instrument. We will confirm the range and quote it — normally the same working day.

Request a quotation