P-5SLPM-D is a gas mass flow meter covering 0 – 5 SLPM, with 1/8" NPT female process connections, 1.0 PSID pressure drop at full scale and an operating pressure range of 11.5 – 160 PSIA.
Request a quotation| Model | P-5SLPM-D |
|---|---|
| Type | Gas mass flow meter |
| Full-scale range | 0 – 5 SLPM |
| Process connection | 1/8" NPT female |
| Pressure drop (air to atmosphere, full scale) | 1.0 PSID |
| Operating pressure | 11.5 – 160 PSIA |
bench-scale duty — laboratory reactors, cell-culture gas overlay, analytical instruments and gas-mixing panels where the working flow sits in the single-digit to twenty-litre band.
The published figure is 1.0 PSID measured with air exhausting to atmosphere at full scale. As a meter there is no control valve in the path; this is the restriction of the laminar element alone, and the upstream supply must hold at least this differential above downstream at full-scale flow.
As a rule of thumb, size the instrument so your working flow falls between roughly 30% and 90% of full scale — well above the bottom of the range, where resolution is poor, and below the top, where the pressure drop figure quoted here applies. Instrument calibration is gas-specific: tell us the gas you are metering and we will confirm whether the published range applies directly or needs a correction.
| Model | Type | Full-scale range | Process connection | Pressure drop | Operating pressure |
|---|---|---|---|---|---|
| P-1000SLPM-D | Gas mass flow meter | 0 – 1000 SLPM | 3/4" NPT female | 6.0 PSID | 11.5 – 160 PSIA |
| P-100SCCM-D | Gas mass flow meter | 0 – 100 SCCM | 1/8" NPT female | 1.0 PSID | 11.5 – 160 PSIA |
| P-100SLPM-D | Gas mass flow meter | 0 – 100 SLPM | 1/4" NPT female | 2.5 PSID | 11.5 – 160 PSIA |
| P-10SLPM-D | Gas mass flow meter | 0 – 10 SLPM | 1/8" NPT female | 1.0 PSID | 11.5 – 160 PSIA |
| P-1SLPM-D | Gas mass flow meter | 0 – 1 SLPM | 1/8" NPT female | 1.0 PSID | 11.5 – 160 PSIA |
| P-200SLPM-D | Gas mass flow meter | 0 – 200 SLPM | — | — | 11.5 – 160 PSIA |
| P-20SLPM-D | Gas mass flow meter | 0 – 20 SLPM | 1/8" NPT female | 1.0 PSID | 11.5 – 160 PSIA |
| P-2SLPM-D | Gas mass flow meter | 0 – 2 SLPM | 1/8" NPT female | 1.0 PSID | 11.5 – 160 PSIA |
| P-500SCCM-D | Gas mass flow meter | 0 – 500 SCCM | 1/8" NPT female | 1.0 PSID | 11.5 – 160 PSIA |
| P-500SLPM-D | Gas mass flow meter | 0 – 500 SLPM | 3/4" NPT female | 4.0 PSID | 11.5 – 160 PSIA |
| P-50SLPM-D | Gas mass flow meter | 0 – 50 SLPM | 1/4" NPT female | 2.0 PSID | 11.5 – 160 PSIA |
See the full P series gas mass flow meters table or compare all 37 ranges.
Send the gas, the flow you need to measure or control, and the pressures either side of the instrument. We will confirm the range and quote it — normally the same working day.
Request a quotation